CRISP Autofokus System

Das CRISP System ist dafür ausgelegt, den Fokus über die Zeit aufrechtzuerhalten, d. H. thermische und andere Faktoren zu kompensieren, die dazu führen können, dass die Probe über die Zeit aus dem Fokus driftet. It also can be used to maintain a given focal point while scanning the sample in XY. If you are looking to find the optimal focal point while scanning through the sample in Z please see our video autofocus system.

Please Note:
CRISP will not work with index matching mounting media. This is how most prepared slides are made. There is no interface with a refractive change near the sample to reflect the light. CRISP relies on reflected light from the sample to detect focus position. Often the reflected light comes from small refractive index discontinuities at sample surfaces. The amount of light reflected at a dielectric interface is given by: R=(n1-n2)^2/(n1+n2)^2
Where n1 and n2 are the reflective indexes of the adjoining dielectric materials. There must be a difference between the refractive indexes in order for CRISP to work.

Theory of Operation

The CRISP system projects the image of a mask illuminated with an IR LED into the sample plane. Only one half of the objective pupil is illuminated. This means that the point spread function of the objective is highly skewed, so that the reflected image of the mask will move laterally as focus is changed. CRISP detects this lateral motion of the LED image to obtain a focus error that is used to close the focus positioning loop.

Installation

CRISP is usually installed with an ASI’s Dual C-Mount Splitter (DCMS) that contains the required dichroic beam combiner and blocking filters and provides the C-Mount port for the camera.

CRISP Drawing 01 01 1030x582 3

YouTube

Mit dem Laden des Videos akzeptieren Sie die Datenschutzerklärung von YouTube.
Mehr erfahren

Video laden

Newsletter

Informieren Sie sich als Erster über alle neuen Produkte in unserem Sortiment und erfahren Sie die aktuellsten Forschungsergebnisse in der Mikroskopie in unserem monatlichen, kostenlosen Newsletter.
Frank Pobig - Geschäftsführer Micrasys e.K. - Closeup

Haben Sie Fragen?

Rufen Sie uns einfach an oder senden Sie uns eine Mail.

× WhatsApp Support